Brookfield DV-I Prime User Manual
Page 19
Brookfield Engineering Laboratories, Inc.
Page 19
Manual No. M/07-022-D0613
NOTE:
If the motor is off or the RPM is 0.0, the maximum viscosity displayed will be 0.0
cP (or 0.0 mPa
.
s).
II.5 CGS or SI Units Selection
Pressing and holding the
AUTO RANGE
key during power on will enable the viscosity display to
be read in either CGS or SI units. To change the unit format:
1. Turn the power off.
2. Press and hold the
AUTO RANGE
key and turn the power ON.
The DV-I PRIME will retain the unit selection when the viscometer is turned OFF.
CGS
SI
Viscosity
cP
mPa
•
s
II.6 Temperature Display in °F or °C
Pressing and holding the
DOWN ARROW
key during power on will enable the temperature display
to be read in either degrees Fahrenheit or degrees Centigrade. To change the units format:
1. Turn the power OFF.
2. Press and hold the
DOWN ARROW
key and turn power ON.
The DV-I PRIME will retain the unit selection when the viscometer is turned OFF.
Figure II-12 depicts the changes to the default screen when displaying temperature in the Fahrenheit
scale and viscosity display in SI units:
mPas123.4 70.1°F
10 RPM
% 89.7
Figure II-12
II.7 Out of Range
Brookfield recommends taking viscosity readings between 10% and 100% of scale. The DV-I
PRIME gives indications for out of specification or out-of-range operation. When % (Torque)
readings exceed 100.0 % (over-range), the display changes to that shown in Figure II-13:
cP EEEE
S01
10 RPM
% EEEE
Figure II-13
You must change either speed or spindle to correct this condition. If you operate at spindle speeds
that produce % (Torque) below 10.0 % (under-range), the DV-I PRIME flashes both % (Torque)
and cP (Viscosity) on and off: