Directions for Use

59

Failed to print

Printer Off-line
Printer out of paper
Printer not connected

Confirm that printer is online.
Confirm printer has paper.
Confirm printer connected.
Using another application, test the printer.
Make sure proper printer is selected with application.

Failure during File transfer

Loss of network connection

User prompted to “Retry” or “Abort”
“Retry” will attempt to reestablish network connection and
continue to transfer file.
“Abort” will lose all unsaved changes and exit the Holter
Application.

Condition

Cause

Action

Table 23. Holter Recorder Acquisition Errors
Recorder LCD displays the error code and Yellow LED (HR100) flashes the error code,
pause for two seconds, and repeats. Remove the battery to repeat diagnostic tests.

Code

LCD and LED

Description

E1

Not used

E2 - Initialize

Configuration error – recorder SD card contains data or awaiting Bluetooth initialization

E3 - BATT

Low/weak battery

E4 - SD Card

Secure digital card not detected

E5 - Call Service

Internal data processing error

E6 - Call Service

Power on self test failed

E7 - Call Service

Wireless data link error

E8 - SD Card

Error writing file to secure digital card

E9 - SD Card

Error reading file from secure digital card