beautypg.com

Operation, To set the time (hr100), To set the time (hr 300 & hr 1200) – Welch Allyn Holter System - User Manual User Manual

Page 16

background image

12

Directions for Use

Welch Allyn Holter System

Operation

To Set the Time (HR100)

Batteries must to stay in the recorder for a minimum of 10-minutes to charge the clock
backup cell.

1.

Remove the battery cover from the recorder (see

Figure 1

on page 11).

2.

Remove the Secure Digital Card from the Secure Digital port located at the rear of the
device (see

Figure 1

on page 11).

3.

Insert one new, AA alkaline (LR 6) batteries into the recorder battery compartment.
Note the polarity.

4.

The LCD window shows the time after the recorder self-diagnostics is complete
(approximately 10 seconds). Time is displayed in 24-hour time format and Date is
shown in Year Month Day format.

5.

Enter time set mode by pressing the Top and Bottom Event switches simultaneously
for 2-seconds.,the hour digit starts blinking

6.

Increase the value shown with the Top Event switch. Use the Bottom Event switch to
toggle between Hours, Minutes, Year, Month and Day.

7.

Press the Top and Bottom Event switches to save time and date shown. Remove the
batteries.

To Set the Time (HR 300 & HR 1200)

Batteries must to stay in the recorder for a minimum of 10-minutes to charge the clock
backup cell.

1.

Remove the battery cover from the recorder (see

Figure 2

on page 11).

2.

Remove the Secure Digital Card from the Secure Digital port located at the rear of the
device (see

Figure 2

on page 11).

3.

Insert two new, AA alkaline (LR 6) batteries into the recorder battery compartment.
Note the polarity.

4.

Slide and secure the battery cover onto the recorder. The LCD window shows the
time after the recorder self-diagnostics is complete (approximately 10 seconds). Time
is displayed in 24-hour time format and Date is shown in Year Month Day format.

5.

Press the function key, the hour digit starts blinking

6.

Increase the hour shown with the up arrow key or decrease the hour shown with the
down arrow key. Use the right and left arrow keys to toggle between Hours, Minutes,
Seconds, Year, Month and Day.

7.

Press the function key to save time and date shown. Remove the batteries.