Troubleshooting – Welch Allyn Holter System - User Manual User Manual
Page 61

Directions for Use
57
Troubleshooting
If the Holter System Application malfunctions, you may be able to resolve the issue with
little loss of time and expense. We suggest following the troubleshooting guidelines
below before returning the unit for servicing.
Contact Welch Allyn Customer Support (see
).
Table 19. Lead Quality Problems
Table 20. Lead Quality Problems Causes and Actions
Table 21. Analysis Errors
Condition
Cause
Actions
General Poor Waveform quality
1, 2, 3, 4, 5
AC interference (even-peaked, regular voltage
superimposed on the waveforms). May resemble or
coincide with muscle-tremor interference.
1, 2, 3, 4, 5
If AC interference persists, turn AC Filter on. See
Muscle tremor interference (random irregular voltage
superimposed on the waveforms). May resemble or
coincide with AC interference
1, 2, 3, 4,
If muscle tremor interference persists, turn Muscle Filter on.
See
Wandering baseline (an upward and downward
fluctuation of the waveform)
1, 2, 3, 4,
If wandering baseline persists, turn baseline filter on. See
Cause
Action
1
Electrodes positioned on a bony area.
Reposition electrodes.
2
Date Code on Hook-up kit expired. Check the expiration date on the
Hook-up kit packaging.
Apply new electrode.
3
Insufficient or dried electrode gel
Apply new electrode.
4
Oily Skin or body lotions
Clean Skin with alcohol or acetone
5
Excessive hair on chest
Shave hair from chest prior to hookup
6
Poor skin preparation.
Remove sensor(s), repeat skin prep, select new position (see
hookup procedure), and apply new electrodes.
7
Faulty patient cable
Replace patient cable.
Condition
Cause
Action
Pause errors
Low Amplitude signal causing misdetection
creating long coupling interval
Pause threshold set too low
Artifact prevents QRS detections creating long
coupling interval.
Reposition electrode to increase signal amplitude.
Check that Pause threshold set to desired threshold.
See
Ventricular errors
Form Incorrectly classified
Excessive noisy trace causes false interpolated
beats
Excessive noisy trace causes distortion of original
QRS signal
Confirm that Form Classifications are correct and edit any
that are not.
Improve patient hookup prep.
Perform beat edit to correct beat classification