beautypg.com

Arrhythmia review – Welch Allyn Holter System - User Manual User Manual

Page 34

background image

30

Directions for Use

Welch Allyn Holter System

Arrhythmia Review

The display for Pause episodes, Ventricular ectopy, and Supraventricular ectopy includes a
histogram of the episodes, a sample 7-second strip (of the worst case occurrence) and a
form associated with the highlighted QRS.

In addition for the Ventricular and Supra Ventricular ectopy a chronological histogram can
be viewed. The Rate tab displays Heart rate graphs for “Beat to Beat” or “HR Minute”.

The R-R Tab displays graphs for “R-R Beat to Beat” or “R-R Histogram”

View any of the events by selecting the respective bar on the histogram and using Next
to view ECG strips for all occurrences.

The individual beats and/or the form may be edited at any time.

Right-click any ECG strip to delete or add to the final report (all beats within the strip are
labeled as artifact or added).

Right-click any ECG strip to print (7-second diagnostic strip will be printed out to the
selected printer).