Directions for Use

55

Recommended Separation Distances Between Portable and Mobile RF Communications Equipment and the Holter recorder

The Holter recorder is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or
user of the Holter recorder can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the Holter recorder as recommended below, according to the maximum output power of the
communications equipment.

Separation Distance According to Frequency of Transmitter (m)

Rated Max. Output

Power of Transmitter

(W)

150 kHz to 80 MHz

d = (1.17)

80 MHz to 800 MHz

d = (1.17)

800 MHz to 2.5 GHz

d = (2.33)

0.01

0.12

0.12

0.23

0.1

0.37

0.37

0.74

1

1.17

1.17

2.33

10

3.70

3.70

7.37

100

11.70

11.70

23.30

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated
using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer.

Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures,
objects, and people.

P

P

P