3 System parameter

In the System Parameters different specific ergometer settings can be carried out.

3.1

Set display

This function allows you to adjust the parameters which can be read out on the display during
exercising. If more then 2 are set to ON they are shown in groups of 2 scrolling. Choosing
default allows you to go to factory settings for the display parameters.
Press the ( ) button to go to the next parameter and switching the actual parameter ON/OFF is
done with ▲ ♥ keys.

3.1.1

Power

Power indication in watts can be switched ON or OFF using the ▲ ♥ keys. The power
displayed is the actual power of the brake. When rpm drops below 25 rpm the load is removed
at once, because the pedal torque becomes very high. Load is put on the brake when rpm is
greater or equal to 30 rpm.

3.1.2

RPM

Swith ON of OFF the Cadence readout in Revolutions per minute. The rpm can always be
read out by the LED bargraph on top of the display independent of this adjustment.

3.1.3

Heartrate

Switch ON or OFF the Heartrate indication in beats per minute. The α symbol is used in the
display.

3.1.4

Torque

Switch ON or OFF the Torque in the readout. The Torque is displayed in Newtonmeter.

3.1.5

Timer

Switches the Timer ON or OFF in the readout. The timer is shown in the display as: → =
mm:ss
. The timer starts when you enter manual, terminal or analog mode and the rpm
increases above 30. The timer is reset when one of these modes is left.

3.1.6

Distance

Switches the distance ON or OFF. The distance is shown in the display as: ▼ = 99,9 km ,
expressing the distance in km. A fictive formula is used (see user guide), holding only for
average cycle conditions and it should not be used for diagnostic purposes.

3.2

Set P-slope

The load regulation steepness is set by this parameter to the default norm-value of 25 watt/sec
or to unlimited.

3.3

Set mode

Set mode is used to define which parameter (like workload, torque, heart rate and linear
constant alpha) has to be adjusted during exercising in MANUAL mode with▲ ♥ keys