The intervals of the measurements should not be too short. Intervals of 2 minutes and more
result in long pauses between the stenoses, so that venous congestion and artiospasm can be
avoided.

D.3

Installing the BP cuff

To install a BP option purchased afterwards, follow the instructions given in the user guide
supplied with the ergometer.
Installation of the BP module itself has to be done by service personnel only.

D.3.1 Analog ECG trigger

The BP module offers the possibility to use an analog heart rate input signal as a trigger for
the BP measurement. Use this option if your ECG device has only an analog output available
instead of TTL. Refer to the user guide for this modification which has to been done by
service personnel only.

D.4

Abort codes BP system


Prior.

Error message

Display

Possible error cause

1


@CHECK ECG
SIGNAL

!

0 / 0 mmHg

ECG trigger not connected.
ECG trigger properties not correct.
ECG trigger polarity not correct.
QRS pulse width smaller than 10 ms.

2


@CUFF
OVERPRESSURE

!

0 / 0 mmHg

Strong arm movements during the blood

pressure measurement.

Air hose is bent or blocked.

3


@AIR LEAK!

0 / 0 mmHg

Cuff is loose or not connected.
Air hose or air-bubble leak.
Air hose not connected.

4


@SERVICE
REQUIRED

!

BP

value

Fatal error ...Get in touch with the company

Lode.

5


@CHECK ECG/
MICROPHONE

!

BP

value

Poor patient hook-up (irregular ECG-signal) or

poor trigger signal.

AND

Abnormal microphone-signal or damaged

microphone.

6


@CHECK
MICROPHONE

!

BP

value

Wrong Microphone placement or defect

microphone.

7


@CHECK ECG/
ARTIFACT

!

BP

value

ECG trigger signal failed during blood pressure

measurement.

Bad or irregular ECG-signal during the blood
pressure measurement.

AND

Excessive movements of the patient during the

blood pressure measurement.

8


@ARTIFACT/ARM

BP

value

Excessive movements of the patient during the

blood pressure measurement.