4 Specifications

4.1

Technical Specifications

Principle

:

eddy current brake

Mains voltage

:

115-230 VAC (adjustable)

Frequency

:

50 Hz, 60 Hz

Patient capacity

:

350 lb - 160 kg

Power consumption

:

120 VA

Workload

:

0-999 W HYP, 0-999 W LIN

Device control

:

manual, external analog, external digital

Accuracy

:

< 3W for P<=100W, < 3% for 100W< P<500W,
<5% for P>500W

Read-out load

:

digital, 0-999 resolution 1 W

Read-out pedaling speed

:

analog, 0-180 rpm or digital 0-255


Output load

:

100, 200 or 500 W 1 volt (max. 999 W)

Output pedaling speed

:

100 min

-1

1 volt

Input voltage for external control

:

100,200 or 500 W 1 volt (max. 999 W)


Safety class Corival and Corival P

:

I type B, IEC 601-1 chapter 18, 19 and 20

Safety class Corival PB

:

IIa type B, IEC 601-1 chapter 18, 19 and 20


Leakage current

:

< 0.4 mA

Isolation resistance

:

> 4 M‘

Earthway resistance

:

< 0.2 ‘


Zero load

:

< 8 W at 60 rpm

Noise level

:

+/- 60dB at workload 200W/60rpm +/-80cm
from crank

4.2

Fuses

115V

:

2 x 1,25A slow

230V

: