beautypg.com

Welch Allyn CP 200 12-Lead Resting Electrocardiograph - User Manual User Manual

Page 39

background image

Directions for Use

Chapter 3 Reviewing the System Settings

35

Transferring a Configuration to Another
Electrocardiograph

You can transfer your configuration from one CP 200 electrocardiograph to another. If you
have the spirometry option, its configuration is included in the transfer.

To Transfer a Configuration

1.

Insert a storage device (SD memory card or removable USB device) into an
electrocardiograph that is configured as desired.

2.

Press the Menu key

.

3.

Choose System Settings > Device Info > Manage Settings > Export Files >
Export Configuration Files.

The files are copied to the storage device.

4.

Remove the storage device, and insert it into another CP 200 electrocardiograph.

5.

Choose System Settings > Device Info > Manage Settings > Import Files >
Import Configuration Files.

The files are copied to the electrocardiograph, and then it reboots.

Figure 24. Configuration File Transfer

Configuration file

or