RECOMMENDED ACTIONS:
Press the

Date/Time

button to access the Set Date and Time menu. Use the

Select

and

Set

buttons to set

the correct date and time.


MESSAGE:

CO

2

cannula or hose occluded

MEANING:
The CO

2

system has detected a flow restriction in its intake.

RECOMMENDED ACTIONS:
Check the tubing for kinks, bends, or blockage.
Check the patient’s nasal cannula for blockage.
Replace the watertrap.


MESSAGE:

CO

2

failure, service required

MEANING:
The Atlas Monitor has detected a malfunction of the CO

2

system.

RECOMMENDED ACTIONS:
A mechanical or electrical fault has been detected in the CO

2

system. Remove the Atlas Monitor from

service. The unit must be repaired.


MESSAGE:

CO

2

Reset Failed

MEANING:
When performing a CO

2

Reset operation, the Atlas Monitor samples CO

2

-free air through the Scrubber, to

use as a baseline reference. This reset operation has failed.
RECOMMENDED ACTIONS:
Turn the Atlas Monitor off and back on, and retry the CO

2

Reset operation. Repeat several times if

necessary. If repeated attempts fail, then the unit must be repaired.


MESSAGE:

CO

2

reset recommended

MEANING:
The Atlas Monitor has determined that the baseline (no CO

2

) setting should be verified by performing a

CO

2

Reset procedure. The Atlas Monitor will recommend a CO

2

Reset operation about once per month.

RECOMMENDED ACTIONS:
At a convenient time, perform a CO

2

Reset procedure using the CO

2

Scrubber included with your model

623xx. The CO

2

system is safe to use in the interim before you perform the reset.