65

Degree of Protection Against Electric Shock

Type CF: Defib Protected

Mode of Operation

Continuous operation

Degree of Protection Against Harmful Ingress of
Water

IPX2 (IEC 60529)

Degree of Safety of Application in the Presence of
a Flammable Anesthetic

Equipment not suitable for use in the presence of a
Flammable Anesthetic Mixture.

Medical Device Directive Classification

II b

FDA Device Classification

II

Canadian Medical Device Directive

III



A.2 SpO

2

- Nellcor

SpO

2

Range

40 to 100%

Measurement Method

Functional saturation

SpO2 Accuracy

Nellcor (no motion)
Nellcor (with motion)

70 to 100%, (Specified at 28º to 42º C sensor
temperature)
± 2 digits
± 3 digits

Pulse Range

20 to 250 beats/minute

Pulse Accuracy
Nellcor (no motion)
Nellcor (with motion)


± 3 digits at 20 to 250 bpm
± 5 digits at 55 to 125 bpm

Plethysmograph

10 segments, vertical bar graph; proportional to
pulse volume

Alarms Low

SpO

2

High and Low Pulse Rate (if ECG not active)

Alarm range

SpO

2

: 60 to 99%

Pulse Rate alarms use ECG Heart Rate settings
when ECG not used

Audible signals (pulse tone)

Frequency tracks SpO

2

level

Used alone, pulse tone tracks SpO

2

pulse

Used with ECG, pulse tone tracks ECG Heart Rate

Audible volume (pulse tone)

8 levels including OFF

Update frequency – Nellcor

Every heart beat

Sensor measurement wavelengths – Nellcor

Red: 660 nm, nominal; Infrared: 890 nm, nominal