beautypg.com

Welch Allyn Atlas Monitor - User Manual User Manual

Page 38

background image

26

suspended alarms will not alert the nurse station, nor
will alarms that occur during a silence period, until
the period is over. See the Technical Service Manual
for details on the Nurse Call relay.

Warning

: it is the user’s responsibility to

implement the interface between the Nurse
Call system and the Welch Allyn Atlas
Monitor. It is also the user’s responsibility to
adequately test the interface between the
Monitor and the Nurse Call system to ensure
that the desired functionality is operational.

Silencing alarms – You can turn off the audible
alarm sound for a brief period by pressing the

Silence

button. The LED will light and no audible

alarms will be heard for 60, 90, or 120 seconds,
depending on the alarm silence duration. When
alarms are silenced, they will not trigger a Print
On Alarm. When the silence period is over, an
alarm that is still active will trigger a Print On
Alarm if the Print On Alarm configuration is set
to Yes.

To set the alarm silence duration, enter Advanced
Configuration by pressing

Date/Time

, then

pressing

Trend

. Use either

Select

button to

highlight Silence duration. Press the

Set

button

to choose 60, 90 or 120 seconds. Exit by pressing
the

Trend

button again.


Pressing the

Silence

button silences all alarms,

including new alarm events that may occur after
you press the button. However, even with no
sound, those measurements that are outside their
limits will still be noticeable, by the flashing
numbers or dashes, or blank displays.

At the end of the silence period, when the LED on
the

Silence

button turns off, any measurement

still outside its limits will cause another audible
alarm.

If you press the

Silence

button during the silence

period, while the bell icon is still lit, the silence
period ends, and any alarm condition, new or old,
will sound the audible alarm.

Suspending alarms – Suspending an alarm

means preventing a patient measurement from
triggering an alarm. Suspended alarms are “off” in
the sense that they will not sound the alarm for a
period of time, but they are “on” in the sense that
the patient’s vital signs are still measured and
displayed. Suspended measurements flash if they
are outside the limits, but they do not cause an
alarm sound. When an alarm is suspended, it will
not trigger a Print On Alarm. When the alarm is
unsuspended, an alarm that is still active will
trigger a Print On Alarm if the Print On Alarm
configuration is set to Yes.

To suspend a measurement alarm, press the
appropriate ALARMS Off button and the red
LED in the center of the button will light. There
are 4 ALARMS Off buttons, and they suspend
the groups of measurements in the following
table:

Suspending Alarms

“Off” button

label

Suspends alarm

HR ALARMS

Heart Rate

CO

2

/RESP

ALARMS

ETCO

2

and Respiration

Rate

SpO

2

ALARMS

Oximetry and Pulse Rate

BP ALARMS

Systolic, Diastolic and
MAP


Removing suspension of alarms – When the
“Francais” language option is selected in the
Advanced Configuration Menu, the alarms can
only be suspended for 3 minutes. To remove
suspension before the 3 minutes have elapsed,
press the

ALARMS Off

button again.


When the Atlas Monitor is set to all other
languages including Canadian French, alarms will
remain suspended until they are turned back on or
the monitor is turned off. To remove suspension,
press the

ALARMS Off

button again. Remember,

suspended alarms do not time out automatically.
Alarms are not suspended at power up, but an
alarm will not sound until valid measurements
have begun.