Atlas Monitor model 623xx showing ETCO

2

waveform display,

controls, and watertrap connector

Setting alarm limits –To set the CO

2

and

Respiration Rate alarm limits, use the

Select

and

Set

buttons on the left side of the monitor to

follow these steps:
• Press the left Select button to cycle through

MAP

HI

and

LO

,

Heart Rate

HI

and

LO

,

Respiration

HI

and

LO

, then CO

2

HI

and

LO

.

• Each push of Select moves you to the next

limit. Stop at the limit you want to set. The
measurement and

HI

or

LO

will flash,

indicating which limit is being changed.

• Press the Set button up or down to raise or

lower the alarm limit. (When you come to
the end of the range, the numbers will stop
changing).

• Press the Select button again to go to the

next limit, or press it several times until
none of the measurements flash and no

HI

or

LO

indicators are lit. The instrument is now

in its normal measurement mode. (If you do
not press any button for 10 seconds, the
instrument will automatically revert to its
normal measurement mode).