78


PROBLEM:
Blood pressure automatic timing stops.
RECOMMENDED ACTIONS:
Safety mechanisms automatically stop the timed blood pressure measurements when the Atlas Monitor
repeatedly fails to make a measurement within 3 minutes

;

Determine the reason that the measurements are failing and restart the timed cycle. Press

BP

Start/Cancel

to manually start a measurement; Auto timing will resume.



E.1.2 CO

2

PROBLEM:
Pump does not start.
RECOMMENDED ACTIONS:
Watertrap not fully inserted.
CO

2

not selected as Second trace selection in Advanced Configuration menu.



PROBLEM:
Display does not start - waveform, Respiration Rate, CO

2

not displayed.

RECOMMENDED ACTIONS:
Watertrap not fully inserted.
CO

2

not selected as Second trace selection in Advanced Configuration menu.



PROBLEM:
Long delay in starting/stabilizing waveform display or numeric displays.
RECOMMENDED ACTIONS:
CO

2

function is specified to begin display within 10 seconds, but not to reach full specification for up to 5

minutes.
Restriction in sampling tubing.
Poor placement of nasal cannula in patient’s nose.
Patient breathing orally.
Patient breathing very shallowly.


PROBLEM:
CO

2

readings inaccurate.

RECOMMENDED ACTIONS:
Pressure or suction being applied to sampling tubing – check for flow interference from ventilation,
oxygen, anesthetic supplies.
High concentration of oxygen or nitrous oxide requires compensation calculation – see section


A.10 End Tidal CO2 (Model 623xx).
Poor placement of nasal cannula in patient’s nose.