E.1.8 Sounds

PROBLEM:
Pulse or alarm sounds are too loud or too quiet when unit is turned on.
RECOMMENDED ACTIONS:
Sound levels can be changed and then set as default values – see section: Managing the Alarms


PROBLEM:
No pulse tone.
RECOMMENDED ACTIONS:
Check default volume setting for pulse tone – see section: Managing the Alarms.
Press SpO

2

volume up button several times.


If SpO

2

is not active and ECG is active:

Pulse tone will not sound until a stable Heart Rate is measured and displayed by ECG – this can
take up to 20 seconds;
If Heart Rate is displayed, have unit repaired.

If SpO

2

is active and ECG is active:

Pulse tone will not sound until a stable heart rate is measured and displayed by ECG – this can
take up to 20 seconds;
If Heart Rate is displayed, have unit repaired.

If SpO

2

is active and ECG is not active:

If Pulse measurement is displayed, have unit repaired.


E.1.9 SpO

2

PROBLEM:
No SpO

2

reading or plethysmograph display.

RECOMMENDED ACTIONS:
Check that the sensor is properly installed in the Atlas monitor.
Check connection of SpO

2

sensor to extension cable.

Check patient: see section Connecting the SpO2 Fingerclip Sensor


E.1.10 Temperature

PROBLEM:
Temperature display is blank
RECOMMENDED ACTIONS:
No temperature probe has been detected: check connection at Atlas Monitor front panel.