12 Appendix C: Calibration and Maintenance

The Atlas Monitor must be serviced by authorized Welch Allyn personnel or agents at 6 month intervals.
Maintenance requirements are specified for 6 month, 12 month and 24 month service intervals. The
monthly CO

2

Reset operation can be performed by the user.


Service Interval

Maintenance Requirements

Every month

CO

2

Reset

Every 6 months

CO

2

Calibration

Button test
LED test
Printer test

Every 12 months

All 6 month maintenance requirements
BP Calibration
ECG Calibration
Battery test

Every 24 months

All 6 month maintenance requirements
All 12 month maintenance requirements
ECG signal gain & noise test
Temperature measurement check with calibrated probe
SpO

2

measurement check and functional tests