beautypg.com

3 monitoring heart rate and the ecg waveform – Welch Allyn Atlas Monitor - User Manual User Manual

Page 27

background image

15

2.3 Monitoring Heart Rate and the ECG Waveform


The trace displays one of the ECG leads. This lead is indicated above the right end of the trace, near the

symbol and heart rate value. Choose the lead with the

Lead Select

button. Heart rate is displayed at

the right.

ECG displays – The ECG waveform is always
displayed on the upper half of the screen. The
ECG waveform cascades (continues) from the
upper half of the screen to also appear at the
bottom portion when other waveforms are not
selected in Advanced Configuration. The Heart
Rate is displayed above the right end of the top
waveform, near the

♥ symbol. The symbol for

the selected ECG lead is shown to the right of
the Heart Rate.

There is always a scale reference bar shown to
the left of the upper ECG waveform. This scale
bar has a height that represents a 1 mV signal.
The apparent height of the scale bar will vary
depending upon the ECG gain setting being
used, but will always correspond to a 1mV
signal.


Selecting leads - Press Lead Select to change
the lead display. The ECG function can use
either 3 wire leads - I, II, III , or 5 wire leads -
I, II, III, aVR, aVL, aVF, and V. 3 wire or 5
wire lead setting is selected in the Advanced
Configuration menu.


Note: When using 3 wire leads, the ECG lead

set must be set correctly. Incorrect results
and noisy waveforms can be obtained if the
system is configured for 5 wire leads when
using 3 wire leads.



Alarms
- When either of the Heart Rate limits
are exceeded, an audible alarm sounds and the
affected measurement flashes. If the Print On
Alarm configuration is set to Yes, this alarm will
trigger an automatic print. You can silence the
Heart Rate alarm, and all alarms, by pressing the

large blue

Silence

button at the right side of the

instrument. This will silence all alarms for 60,
90, or 120 seconds, depending on the setting
selected in Advanced Configuration. However,
any measurement still outside the set limits will
flash. When alarms are silenced, they do not
trigger a Print On Alarm.

To suspend the Heart Rate alarms, press the

HR

ALARMS Off

button so the red LED in the

button lights. A suspended alarm will still flash
if it goes outside the range of the limits, but it
will not sound the audible alarm. When an alarm
is suspended, it will not trigger a Print On
Alarm.

Note:
The Heart Rate alarm can be suspended

for only 3 minutes when the language setting
is configured to “Francais”. If the language
is set to any other language, pressing the

HR

ALARMS Off

button will suspend the alarm

until you remove the suspension.


If ECG is inactive for any reason, the Heart Rate
display will be dashes “---” and the Heart Rate
alarm will respond to the pulse oximetry rate.


Setting alarm limits –To set the Heart Rate
alarm limits, use the

Select

and

Set

buttons on

the left side of the monitor to follow these steps:
• Press the left Select button to cycle through

MAP

HI

and

LO

,

Heart Rate

HI

and

LO

,

Respiration

HI

and

LO

(models 622xx or

623xx)

, then (model 623xx) CO

2

HI

and

LO

.

• Each push of Select moves you to the next

limit. Stop at the limit you want to set. The
measurement and

HI

or

LO

will flash,

indicating which limit is being changed.

• Press the Set button up or down to raise or