7

CO

2

/ RESP

ALARMS

Trend

Lead Select

Print

AC

Select

Set

Off

Off

Select

Set

PG 2-5

Blood

Pressure

Connector

Atlas Monitor showing NIBP displays, controls and tubing

connector

Setting alarm limits - To set the Systolic and
Diastolic alarm limits, use the

Select

and

Set

buttons on the right side of the monitor to follow
these steps:
• Press the

Select

button to cycle through

SpO

2

LO

, then Systolic

HI

and

LO

, and

Diastolic

HI

and

LO

.

• Each push of

Select

moves you to the next

limit. Stop at the limit you want to set. The
measurement and the

HI

or

LO

LED will

flash, indicating which limit is being
changed.

• Press the

Set

button up or down to raise or

lower the alarm limit. (When you come to
the end of the range, the numbers will stop
changing).

• Press the

Select

button again to go to the

next limit, or press it several times until
none of the measurements flash and no

HI

or

LO

LEDs are lit. The instrument is now in

its normal measurement mode. (If you do
not press any button for 10 seconds, the
instrument will automatically revert to its
normal measurement mode).

Note: If the language setting is configured to
Chinese and the alternate BP field to kPa, the
blood pressure will be displayed in kPa on the
screen, but the Systolic and Diastolic readings
on the right side of the monitor will be displayed
in mmHg. When setting the alarm limits, the

HI

and

LO

alarm limits will be displayed in mmHg

on the right side of the monitor and in kPa on the
screen.