Welch Allyn Atlas Monitor - User Manual User Manual
Page 19

7
CO
2
/ RESP
ALARMS
Trend
Lead Select
AC
Select
Set
Off
Off
Select
Set
PG 2-5
Blood
Pressure
Connector
Atlas Monitor showing NIBP displays, controls and tubing
connector
Setting alarm limits - To set the Systolic and
Diastolic alarm limits, use the
Select
and
Set
buttons on the right side of the monitor to follow
these steps:
• Press the
Select
button to cycle through
SpO
2
LO
, then Systolic
HI
and
LO
, and
Diastolic
HI
and
LO
.
• Each push of
Select
moves you to the next
limit. Stop at the limit you want to set. The
measurement and the
HI
or
LO
LED will
flash, indicating which limit is being
changed.
• Press the
Set
button up or down to raise or
lower the alarm limit. (When you come to
the end of the range, the numbers will stop
changing).
• Press the
Select
button again to go to the
next limit, or press it several times until
none of the measurements flash and no
HI
or
LO
LEDs are lit. The instrument is now in
its normal measurement mode. (If you do
not press any button for 10 seconds, the
instrument will automatically revert to its
normal measurement mode).
Note: If the language setting is configured to
Chinese and the alternate BP field to kPa, the
blood pressure will be displayed in kPa on the
screen, but the Systolic and Diastolic readings
on the right side of the monitor will be displayed
in mmHg. When setting the alarm limits, the
HI
and
LO
alarm limits will be displayed in mmHg
on the right side of the monitor and in kPa on the
screen.