Appendix E: Configurations and test protocols

Note

The protocols Included are DP 4s (default) and DP 2s.

Name

No. of

Frequenci

es

F2 Frequency

L1/L2

Maximum Average

Time

Pass

SNR

# Passing

Frequency for

Test Pass

DP 4s

4

2.0, 3.0, 4.0, 5.0 kHz 65/55

4 seconds

6 dB

3

DP 2s

4

65/55

2 seconds

6 dB

3

Note

The DPOAE protocols differ in the amount of averaging time per frequency.
The DP 4s protocol is the default. This averaging time is the time the
screener uses to render a decision on that frequency. The DP 2s (2 second)
protocol uses 2 seconds per frequency and results in faster test times at a
minimum of 8 seconds. However, it may increase the false referral rate.
The DP 4s (4 second) protocol uses 4 seconds averaging time per
frequency and results in slower test time at a minimum of 16 seconds per
tes. However, it may reduce the false referral rate.