Status/Error messages
Display messages
Message
Definition
Attach Probe
No probe is detected at the start of a test.
Device not Responding
The printer is not responding to queries from the instrument.
BT Device Not Found.
The paired Bluetooth
®
device cannot be detected by the
screener. The screener may be turned off or too far away.
BT Error #xxx.
There is an error condition with the Bluetooth
®
device. Check
the status.
BT Not Configured
The screener is not paired with any Bluetooth
®
device.
Fit Error Cannot Obtain P
For a DP test, the desired intensity level (L1 or L2) cannot be
obtained within allowable limits. Refit the eartip and probe and
attempt the test.
Fit Error Too High
The level of the calibration tone is too high. Refit the eartip and
probe and retry the test.
Fit Error Too Low
The level of the calibration tone is too low. Refit the eartip and
probe and retry the test.
Limit Error
Overflow error during the calculation of the DFTs for a DP test.
Repeat the test.
Memory Almost Full
Saved tests are within 5 tests of the 250 maximum limit.
Memory Full!
The maximum saved test limit of 250 is reached. Clear the
memory before any additional tests can be performed.
Power Low!
The battery charge level is too low for operation. Charge the
battery in order to administer additional tests.
Printer Error
There is a problem with the printer. Check the printer status.
Printer Paper Out!
The printer paper has run out. Load printer paper.
Time/Date Error
The clock is checked during power on to ensure it has not lost
time and been reset. In the case of clock reset, this message is
shown. Set the correct date/time.