Manage results

Results storage and reporting

When the screener is set in its default settings, the instrument stores the results from

one patient (left and right ears) in its non-volatile memory for subsequent printing. The

screener is capable of storing up to 250 test results.

The results are displayed on the LCD on the front of the device and are stored in the

device’s internal memory. After testing is complete, results can be printed using the

printer or exported to a computer.

Save/Store test results mode

The screener automatically saves the results of completed tests in the non-volatile

memory even if the battery is temporarily discharged. However, the screener is not

intended for long-term storage of test results.

Note

You are strongly encouraged to print or transfer all test results at the
completion of testing to avoid potential loss of data.

The screener has two operating modes for saving and storing test results:

SAVE L/R

SAVE 250

SAVE/LR is the default operating mode.

SAVE L/R mode

When operating in the default SAVE L/R mode, the screener saves the most recent test

results for each ear and prints or transfers only these results. This allows you to retest a

patient after a REFER result and to print or transfer only the most recent test result for

each ear.

Welch Allyn recommends that you print the results after each patient when using the

default mode.

SAVE 250 mode

When operating in the SAVE 250 mode, the screener can save up to 250 tests. There are

two methods of using the SAVE 250 mode: