Ear canal volume

Explanation

Blue

The ear canal volume is too large for the test to begin. The probe is not in the ear or

there is a large leak.

Green

The ear canal volume is in the target range for testing. The test will begin automatically

if the probe fit is stable.

Orange

The ear canal volume is too small for the test begin. The eartip may be pressed against

the ear canal wall or the probe tube may be completely clogged.

Troubleshoot the AutoStart probe check

Appropriate adjustments of the eartip position and size selection should be made until

the indicator falls within the green area and remains stable. If the test will not progress

past the AutoStart probe check stage:

1. Change the probe tube.
2. Confirm that the probe connector is fully seated in the socket.
3. Run the AutoStart probe check again.

30 Operating instructions

Welch Allyn

®

OAE Hearing Screener