Specifications

Welch Allyn MicroTymp3 portable tympanometric instrument

Technical specifications

Probe tone

226-Hz Amplitude: 85 ± 3 dB SPL

1000-Hz Amplitude: 83 ± 3 dB SPL

Frequency Accuracy: ± 2%

Total Harmonic Distortion: 3% maximum

Signal Type: Continuous Sinusoid

All specifications measured in an ANSI HA-1 (2.0 cc) coupler.

Pressure measurement system

Direction of sweep: positive to negative pressure

Sweep Rate: 400 daPa/sec average during data acquisition period (applies only at sea level)

Range: +200 to -400 daPa

Display resolution: 20 daPa

Accuracy: ±15% or ±10 daPa, whichever is greater

Compensation: Auto-zero every test cycle

Admittance measurement system

226-Hz mode

Range: 0.2 to 4.0 mmho total

0.0 to 1.5 mmho for Ya

0.2 to 2.5 cc for +200 Vea

Accuracy: ±0.1 mmho or ±5%, whichever is greater

Positive tail compensation

1000-Hz mode

Range: 0.9 to 16.6 mmho total

0.0 to 6.5 mmho for Ya (for ear canal volumes

2.0 cc)

0.2 to 2.5 cc for +200 Vea

Accuracy: ± 0.4 mmho or ± 5%, whichever is greater

Negative tail compensation