Maintenance and troubleshooting
Welch Allyn MicroTymp3 portable tympanometric instrument
Recycle the Battery
Check with your local waste management agency for the proper recycling procedure for a
Lithium-ion battery.
Replace the tips
Replace the probe tips after six months of use.
Replace the paper
The Printer/Charger signals the need for changing the paper in one of two ways:
•
A pink strip appears along the edge of the paper indicating the paper is nearing the
end of the roll.
•
The POWER indicator flashes in single pulses indicating that there is no paper, and no
printing can occur.
1.
Remove the paper access cover by pulling up on the front edge.
Lithium-ion battery.
Must be recycled or
disposed of properly.
Note
•
Use only an appropriate heat-sensitive paper or the Printer/Charger life may
be shortened and the warranty voided.
•
The paper is thermally activated, so it must be stored in a cool, dark location
to prevent exposure and degraded performance.
•
Because the paper is thermally activated, no printing will appear on the paper
if it is inserted backwards.
•
Do not use transparent adhesive tape on printed portions of the printout, as
those portions will then fade.
WARNING Electric shock hazard. Do not attempt to disassemble the Printer/
Charger. Refer all servicing to Welch Allyn or a Welch allyn authorized service
representative.