beautypg.com

Block, Leak, Block leak – Welch Allyn MicroTymp 3 portable tympanometric instrument - User Manual User Manual

Page 25

background image

Directions for use

Understanding the LCD messages

25

Block

The test cannot continue since the measured admittance is less than 0.2 mmho.

Possible causes:

Probe tip is lodged against canal wall

Ear canal occluded

Collapsed ear canal

Extremely small ear

Probe tip is clogged with cerumen

Possible solutions:

Reposition the probe tip

Perform otoscopy to check for occlusion

Remove cerumen from probe tip

Leak

The test cannot proceed since desired pressures within the ear have not been achieved.

Possible causes:

Probe tip is not completely sealed in the ear canal

Excessive movement of patient or practitioner

Probe tip dislodged during a test

-400

-200

0

200

2

0.5

0

1.0

1

GR

A

GR

C

PRESSURE - daPa

Ya - mmho

+200 Vea - cc