Understanding the LCD messages

Welch Allyn MicroTymp3 portable tympanometric instrument

Open

The test has not begun since a valid ear cavity has not been detected.

Possible causes:

Instrument is not in an ear

Probe tip is not completely sealed in an ear

Perforated tympanic membrane

Ear with patent tympanostomy tubes

Extremely large canal (> 2.5 cc)

Possible solutions:

Reposition the probe tip

Perform otoscopy to check for perforated membrane or patent tympanostomy tubes

Test

The TEST message indicates that the test has started. Test data will begin to appear
immediately.