Interpreting test results

Welch Allyn MicroTymp3 portable tympanometric instrument

Normative values

The table below presents normative values for 1000-Hz tympanometry for NICU babies
and full-term 2- to 4-week-old babies. [Margolis, R.H., Bass-Ringdahl, S., Hanks, W.D.,
Holte, K., Zapala, D.A. Tympanometry in Newborn Infants - 1 KHz Norms. J. Amer. Acad.
Audiol., 14, 383-392, 2003.]

Interpreting 1000-Hz results

Pass

A normal tympanogram will yield a PASS result. The displayed numerical value represents
the compensated admittance.

Condition which causes a PASS message:

Peak Ya greater than or equal to 0.6 mmho

Normative tympanometric values (1000 Hz)

Tympanometric Measurement

NICU
90% Range

Full-term
90% Range

Peak Ya
(Negative Tail Compensation)

0.6 - 2.7 mmho

0.6 - 4.3 mmho

Tympanometric
Peak Pressure (TPP)

-93 - +53 daPa

-133 - +113 daPa

-400

-200

0

200

2

0.5

0

1.0

1

GR

A

GR

C

PRESSURE - daPa

Ya - mmho

+200 Vea - cc