Directions for use

Completing a test

21

3.

Remove the Handle cover by lifting the bottom of the cover away from the probe tip.

4.

With a sharp, non-conductive object (like a toothpick or the wooden end of a Q-tip),
move switch 1 to the desired position as shown below.

5.

Replace the cover by inserting the tip end under the front cover, lowering the opposite
end, and replacing the screw.

6.

Tighten the screw in a clockwise direction. To avoid stripping the screw threads, do
not tighten excessively.

72910

REF

+

Lithium

Caution Potential damage to Handle. Do not use the screwdriver provided or
any other metal object to move the switch.

ON CTS

ON

OFF

1 2 3 4

ON CTS

ON

OFF

1 2 3 4

72910

REF

+

-

Lithium

Ion

ON = 226 Hz
OFF = 1000 Hz