Understanding the LCD messages

Welch Allyn MicroTymp3 portable tympanometric instrument

Possible solutions:

Reposition probe tip

Patient and practitioner must remain still

Use a different size probe tip

Increase pressure against the ear

If a leak condition occurs after -100 daPa pressure is reached at 226 Hz, results will remain
on the display. If an identifiable peak is displayed at 226 Hz, you do not have to repeat the
test. If no peak can be identified, repeat the test and try repositioning the tip, using a
different size tip, or increasing the pressure against the ear canal opening.

If a leak condition occurs after -300 daPa pressure is reached at 1000 Hz, the PASS,
REFER, or RETEST results will be displayed.

Right store/Left store

Test results have been stored in the designated memory location. Immediately following
this message, the newly-stored tympanogram reappears along with the right or left
symbol. For instructions on storing and erasing data, see Chapter 2.

Right clear/Left clear

The designated right or left memory location contains no data. Either no data has been
stored or previously stored data has been erased. See Ch. 2 for information on storing and
erasing data.

-400

-200

0

200

2

0.5

0

1.0

1

GR

A

GR

C

PRESSURE - daPa

Ya - mmho

+200 Vea - cc

-400

-200

0

200

2

0.5

0

1.0

1

GR

A

GR

C

PRESSURE - daPa

Ya - mmho

+200 Vea - cc

-400

-200

0

200

2

0.5

0

1.0

1

GR

A

GR

C

PRESSURE - daPa

Ya - mmho

+200 Vea - cc

-400

-200

0

200

2

0.5

0

1.0

1

GR

A

GR

C

PRESSURE - daPa

Ya - mmho

+200 Vea - cc