Model-specific data – Rice Lake Secura - Quintix and Practum Series - User Manual User Manual
Page 95
95
Specifications
User Manual for Secura, Quintix, Practum
Model-specific Data
Secura
®
Model
224
124
513
313
213
5102
3102
2102
1102
Weighing capacity
g
220
120
510
310
210
5100
3100
2100
1100
Readability
mg
0.1
0.1
1
1
1
10
10
10
10
Repeatability (standard deviation)
mg
0.1
0.1
1
1
1
10
10
10
10
Linearity deviation
mg
0.2
0.2
2
2
2
20
20
20
20
Minimum weight as per USP,
typical
g
0.12
0.12
1.5
1.5
1.5
12
12
12
12
Sensitivity drift
between +10 and +30°C
± ppm/K 1.5
1.5
3
3
3
3
3
3
3
Typical stabilization time
s
2
2
1.5
1.5
1.5
1.5
1.5
1.5
1.5
isoCAL:
– Temperature change
– Time interval
K
h
1.5
4
1.5
4
2
6
4
12
4
12
2
6
4
12
4
12
4
24
Display result (depending
on the set filter level)
s
0.2
0.2
0.1/0.2
0.1/0.2
0.1/0.2
0.1/0.2
0.1/0.2
0.1/0.2
0.1/0.2
Weighing pan size
mm
90
90
120
120
120
180
180
180
180
Weighing chamber height
mm
209
209
209
209
209
–
–
–
–
Net weight, approx.
kg
5.1
5.1
5.1
5.1
5.1
5.2
5.2
5.2
5.2
Quintix
®
Model
224
124
513
313
213
5102
3102
2102
1102
Weighing capacity
g
220
120
510
310
210
5100
3100
2100
1100
Readability
mg
0.1
0.1
1
1
1
10
10
10
10
Repeatability (standard deviation)
mg
0.1
0.1
1
1
1
10
10
10
10
Linearity deviation
mg
0.2
0.2
2
2
2
20
20
30
30
Sensitivity drift
between +10 and +30°C
± ppm/K 1.5
1.5
3
3
3
3
3
5
5
Typical stabilization time
s
3
2
1.5
1.5
1.5
1.5
1.5
2
2
isoCAL:
– Temperature change
– Time interval
K
h
1.5
4
1.5
4
2
6
4
12
4
12
2
6
4
12
4
12
4
24
Display result (depending
on the set filter level)
s
0.2
0.2
0.1/0.2
0.1/0.2
0.1/0.2
0.1/0.2
0.1/0.2
0.1/0.2
0.1/0.2
Weighing pan size
mm
90
90
120
120
120
180
180
180
180
Weighing chamber height
mm
209
209
209
209
209
–
–
–
–
Net weight, approx.
kg
4.9
4.9
4.9
4.9
4.9
5.2
5.2
4.7
4.7