System Settings (Setup Menu)
User Manual for Secura, Quintix, Practum
Device
Information
Device information
Manufacturer: Sartorius
Model: SECURA2102
Serial number:
0012345678
Version BAC:
00-50-02.xx CN:1701
Version APC:
01-70-02.xx CN:6587
The manufacturer, model, serial number, and software version of your balance are
displayed here.
Calibration/Adjustment
Calibration/Adjustment
isoCAL
Calibration report
With
Secura
®
and Quintix
®
models, saved reports about the most recent calibration
procedures, including the identified deviations, can be viewed and output to a PC
or Sartorius lab printer.
With Secura
®
models it is also possible to set how the automatic calibration/
adjustment function isoCAL is started.
isoCAL
menu option (Secura
®
only):
t
Select the desired start option for isoCAL.
–
Off
: isoCAL is switched off.
–
Info, manual start
: A note appears on the display when predefined time
intervals or temperature values are exceeded. isoCAL can be started
manually.
–
Automatic
: isoCAL is switched on. Calibration/adjustment starts
automatically when predefined time intervals or temperature values are
exceeded.
Calibration report
menu option (Secura
®
and Quintix
®
only):
t
Select the desired report by date.
If there is more than one calibration report per date, use the
<
and
>
buttons to
navigate through the reports.
t
Select
a
to print the displayed report.
Weighing
Weighing
SQmin
Application
Stability signal
Ambient conditions
Safety level
A number of basic settings for the weighing functions can be defined here. Options
are limited for verified balances. All settings given in the menu are permitted.
SQmin
menu option (Secura
®
only):
If the SQmin function is set up on the balance, the function can be switched
On
or
Off
here (see “SQmin Minimum Weighing (Secura
®
only)" on page 37).
Safety level
menu option (Secura
®
only):
t
Select the desired safety level for
SQmin
,
isoCAL
, and
LEVEL
–
Low
: Information about display elements given in red. Weighing is still
possible.
–
Standard
: As with
Low
, but the weight value is also displayed in gray and
the warning symbol is shown. The weight values are marked as not safe.
A warning appears after 60 seconds if the situation is ignored.
–
High
(factory setting): As with
Standard
, but the weight value in the
applications cannot be printed out or saved if the value is smaller than the
specified minimum initial weight (SQmin).
Calibration report
2012-10-04: 2 Protocols
2012-10-01: 2 Protocols
2012-09-25: 3 Protocols
2012-09-24: 3 Protocols
2012 09 20: 4 Protocols