Error messages – Rice Lake Sartorius Midrics 1 and 2 Series - Operating Instructions User Manual
Page 89

89
Error messages are shown in the main display. Error messages are shown permanently; Information messages are shown for 2 seconds.
After this the program automatically returns to the weighing mode.
Display
Cause Remedy
ERR101
Key is stuck
Release key or
Key pressed whilst switching on
Contact your customer service partner
ERR 320
Operating program memory faulty
Contact your customer service partner
ERR 340
Incorrect operating parameter (EEPROM)
Switch scale off and then on again,
if Err 340 is permanently displayed:
Contact your customer service partner
ERR 341
Loss of data
Contact your customer service partner
ERR 343
Loss of data in storage for
Contact your customer service partner
external Alibi memory transaction number
Inf 01
Data output not compatible with
Set output format correctly
output format
Inf 02
Calibration condition was not met
Do not carry out calibration until after 0 display
e.g. not tared or
Unload the weighing equipment
weighing pan loaded
Tare with the ) key
Inf 03
Calibration could not be completed
Keep to warm-up time and
within a certain time
calibrate again
Inf 07
Last operating function is not permissible
For setting changes please
use in legal metrology
Contact your customer service partner
Inf 08
Load on the scale is too heavy for scale to be
Please check if point 1.12 “Zero-setting
zeroed.
range” was adhered to in your configuration.
Inf 09
Taring not possible for gross zero
Zero the scale
Inf 10
Taring not possible with occupied tare memory
Taring only possible after deletion of the
application program
Inf 22
Reference takeover error,
Place a heavier weight on the platform
weight too small
Inf 23
Error in initializing an application
Contact your customer service partner
Inf 29
Minimum load not achieved
Reduce minimum load
(under Application, menu item 3.6)
Inf 71
Saving of measured value (or input) not possible
None
(e.g. control limit too small or too large)
Inf 72
Saving of measured value not possible
None
(e.g. maximum for item counter reached)
Inf 73
Stored data is deleted or cannot be read
Contact your customer service partner
Inf 74
Function is locked
None
(e.g. menu is locked)
Inf 98
No weighing platform connected
Contact your customer service partner
Inf 99
No weighing platform connected
Contact your customer service partner
NO WP
No weighing platform connected
Connecting weighing platform