Rice Lake Sartorius Midrics 1 and 2 Series - Operating Instructions User Manual
Page 77
Application Parameters: Totalizing
3.5.
Minimum load for automatic
taring and automatic printing
3.5.1*
1 digit
3.5.2
2 digits
3.5.3
5 digits
3.5.4
10 digits
3.5.5
20 digits
3.5.6
50 digits
3.5.7
100 digits
3.5.8
200 digits
3.5.9
500 digits
3.5.10
1000 digits
3.6.
Minimum load for automatically
saving/transferring values
3.6.1*
1 digit
3.6.2
2 digits
3.6.3
5 digits
3.6.4
10 digits
3.6.5
20 digits
3.6.6
50 digits
3.6.7
100 digits
3.6.8
200 digits
3.6.9
500 digits
3.6.10
1000 digits
3.7.
Automatic taring:
first weight tared
3.7.1*
Off
3.7.2
On
3.8.
Start application and load most
recent application data when the
Midrics is switched on
3.8.1
Automatic (on)
3.8.2*
Manual (off)
3.16. Values saved automatically
3.16.1*
Off
3.16.2
On
3.17. Automatic individual or compo-
nent printout when value stored
3.17.1
Off
3.17.2*
Generate printout with
complete standard
configuration each time
O is pressed
3.17.3
Generate printout with
complete standard
configuration only the
first time O is pressed
3.23. Value(s) to be saved
3.23.1*
Net
3.23.2
Calculated
3.23.3
Net and calculated
3.25. Tare function
3.25.1*
Add input value (weight
value) for taring
3.25.2
Tare value can be over-
written
* = Factory setting
§ Press ) to save your settings and (
(repeatedly) to exit the operating menu.
Minimum Load
The minimum load required for automatic
taring of the container weight on the
platform (“autotare first weight") is config-
ured in the operating menu under:
Appl
L
:
3.5.
The minimum amount that a component
must weigh before it can be saved in
totalizing memory is configured in the
operating menu under:
Appl
L
:
3.6.
You can choose from 10 settings,
ranging from
1 digit
to
1000 digits
Example: If the scale interval (d) is 1000 g
and the minimum load is set to 1000 digits
(=1000 scale intervals), a load of at least
1000 g is required for autotaring.
Data Record or Printout
In the operating menu, under:
Appl
L
:
3.17.
you can configure whether a printout
is generated manually, by pressing p,
or automatically when a weight value
is saved in the totalizing memory.
If you select
3.17.1
for this setting,
printouts can be generated only manually,
by pressing p (individual printout).
If you select
3.17.2
(printout of a compo-
nent on request), the component printout
is generated.
The total data record is printed
when you clear the totalizing memory
(by pressing the c key).
77