Weigh with varying tare values – Rice Lake Sartorius Midrics 1 and 2 Series - Operating Instructions User Manual
Page 52

e
1. Switch on the scale
The automatic self-test runs.
Once a readout is shown,
the Midrics is automatically
zeroed and ready to operate.
Press ( to reset the
unloaded weighing platform
to zero at any time.
2. Place empty container on
the platform
)
3. Tare the instrument
Note: If the automatic tare
function is active, you do not
need to press ) to tare the
scale; the tare weight is stored
automatically when you place
the container on the platform.
Wait until a zero value is
displayed together with the
NET
symbol.
4. Place the sample in its
packaging (second tare value)
in the container.
250
5. Enter the tare weight of the
packaging in the current
weight unit using the keypad
(in this example, 250 g).
)
6. Save the package weight
you entered (the two tare
values are added together).
Read the net weight
p
7. Print the results.
G#
+
6.433 kg
T
+
4.183 kg
PT2
+
0.250 kg
N
+
2.000 kg
--------------------
0
8. Clear the tare memory:
Enter a zero (“0") using the
keypad
)
9. Save the value entered (0).
This deletes tare values;
the display shows the gross
value.
p
10. Print the results.
G#
+
6.433 kg
T
+
0.000 kg
N
+
6.433 kg
--------------------
52
Example with Midrics 2:
Weigh with varying tare values; print the results; delete tare values