Key ) - > 2 sec Function
Allocation
Purpose
The key ) - > 2 sec is usually used
for the calibration/adjustment function.
The following additional functions can
be allocated to the key when the service
mode is activated:
– External linearization with default
weights (menu item 1.9.6)
– External linearization with the lineariza-
tion weights (menu item 1.9.7) entered
under menu item 1.18
– Setting the preload (menu item 1.9.8)
– Clearing the preload (menu item 1.9.9)
!
Once linearization has been completed,
or after a preload has been set or
cleared the function of the key )- > 2
sec must be reallocated back to its nor-
mal function in the Setup menu (e.g.
external calibration/adjustment with
default weights)
17
17
Menu structure for key ) - > 2 sec function allocation
1. 9.
Calibration, adjustment
1. 9. 1
External calibration/adjustment with default weights (service mode not required)
1. 9. 3
External calibration/adjustment with user-defined weights (entered under 1-18), (service mode not required)
1. 9. 6
External linearization with default weights
1. 9. 7
External linearization with user-defined weights (entered under 1-18)
1. 9. 8
Set preload
1. 9. 9
Clear preload
1. 9. 10
Key blocked