Example 1:
Enter or change values for standard configuration in single range
mode in the unit set under 1.7.x.
Select menu item ADC CON
)
Confirm menu item ADC CON
(if necessary k)
to select the menu item STAND
Standard configuration
)
Confirm menu item STAND.
Range selection
)
Confirm menu item RANGES
(if necessary press k repeatedly) Select menu item SINGLE
Single range mode
(
Confirm menu item SINGLE
Weights
)
Confirm menu item WEIGHTS
Scale interval
)
Confirm menu item D
( ) k p
Enter a value (e.g. 0.002 kg)
(if necessary press ) repeatedly) until display D appears
k
Select menu item MAX.
Maximum capacity
)
Confirm menu item MAX.
( ) k p
Enter a value (e.g. 30 kg)
(if necessary press ) repeatedly) until display MAX appears
(
Menu item UNITS is displayed
(if necessary press the ) key)
to select available
weighing units (UNITS)
k
Menu item SAVE is displayed
) k )
Save the entered value (YES)
or do not save (NO)
Save
max.
d
Weights
SINGle
Ranges
Stand.
ADC-CON
Example 2:
Enter or change values for standard configuration in a multi-
interval scale in the unit set under 1.7.x.
(the same applies for multiple range mode).
Select menu item ADC CON
)
Confirm menu item ADC CON
(if necessary press
and select the menu item STAND.
) k p)
Standard configuration
)
Confirm menu item STAND.
Range selection
)
Confirm menu item RANGES
(if necessary press k repeatedly) Select menu item MULT. INT
Multi-interval scale
(
Confirm menu item MULT. INT
Weights
)
Confirm menu item WEIGHTS
Scale interval (e.g. 0.002 kg)
)
Confirm menu item D
( ) k p
Enter a value (e.g. 0.002 kg)
(if necessary press ) repeatedly)
until display D appears
k
Select menu item RANGE 1
Enter values for the following in
the same way:
Range limit 1 (e.g. 6 kg)
Range limit 2 (e.g. 15 kg)
Maximum capacity (e.g. 30 kg)
Continue as shown in example 1
after entering the maximum
capacity
max.
Range 2
Range 1
d
Weights
Mult.Int
Ranges
Stand.
ADC-CON
15