CLS-420 Installation Manual
5.2.9
Analog Output Menu
The ALGOUT menu is shown only if the analog output option is installed. If the analog output option is installed,
configure all other indicator functions and calibrate the indicator itself before configuring the analog output.
XXXXXXX
XXXXXXX
ALGOUT
DIGIN
SETPNT
XXXXXXX
PROGRM
PFORMT
SERIAL
CALIBR
XXXXXXX
CONFIG
FORMAT
SOURCE
GROSS
0%
20%
OFFSET
ERRACT
FULLSC
HOLD
000000
number
MIN
10000
number
MAX
TWZERO
NET
TWSPAN
ZEROSC
VERS
MINNEG
OFF
ON
MAXNEG
OFF
ON
Figure 5-13. Analog Output Menu
Table 5-10. Analog Output Menu Parameters
ALG OUT Menu
Parameter
Choices
Description
Level 2 submenus
SOURCE
GROSS
NET
Specifies the scale tracked by the analog output.
OFFSET
0%
20%
Zero offset. Select 0% for 0–10 V or 0–20 mA output; select 20% for 4–20 mA output. This parameter
must be set before calibrating the analog output.
ERRACT
FULLSC
HOLD
ZEROSC
Error action. Specifies how the analog output responds to system error conditions.
Possible values are:
FULLSC: Set to full value (10 V or 20 mA)
HOLD: Hold current value
ZEROSC: Set to zero value (0 V or 4 mA)
MIN
000000
0–9999999
Specifies the minimum weight value tracked by the analog output. Specify a value in the range
0–9999999.
MINNEG
OFF
ON
Turn ON or OFF.
MAX
10000
0–9999999
Specifies the maximum weight value tracked by the analog output. Specify a value in the range
0–9999999
MAXNEG
OFF
ON
Turn ON or OFF.
TWZERO
_
Tweak zero. Enter tweak value to adjust the analog output zero calibration. Use a multimeter to monitor
the analog output value.
TWSPAN
_
Tweak span. Enter tweak value to adjust the analog output span calibration. Use a multimeter to monitor
the analog output value.