5 program menu, Program menu, Table 3-6. program menu parameters – Rice Lake Intrinsically Safe User Manual
Page 34

28
320IS Installation Manual
3.3.5
Program Menu
Figure 3-5. Program Menu
PROGRM Menu
Parameter
Choices
Description
Level 2 submenus
PWRUPM
GO
DELAY
Power up mode. In GO mode, the indicator goes into operation immediately after a brief power
up display test.
In DELAY mode, the indicator performs a power up display test, then enters a 30–second
warm up period. If no motion is detected during the warm up period, the indicator becomes
operational when the warm up period ends; if motion is detected, the delay timer is reset and
the warm up period is repeated.
COUNT
ACCESS
SPLSIZ
Specifies whether operator has access to piece count mode and the default sample size used
for parts counting. See Level three submenu for parameter descriptions.
REGULA
NTEP
OIML
CANADA
NONE
Regulatory mode. Specifies the regulatory agency having jurisdiction over the scale site.
• OIML, NTEP, and CANADA modes allow a tare to be acquired at any weight greater than
zero. NONE allows tares to be acquired at any weight value.
• OIML, NTEP, and CANADA modes allow a tare to be cleared only if the gross weight is at
no load. NONE allows tares to be cleared at any weight value.
• NTEP and OIML modes allow a new tare to be acquired even if a tare is already present.
In CANADA mode, the previous tare must be cleared before a new tare can be acquired.
• NONE, NTEP and CANADA modes allow the scale to be zeroed in either gross or net
mode as long as the current weight is within the specified ZRANGE. In OIML mode, the
scale must be in gross mode before it can be zeroed; pressing the ZERO key in net mode
clears the tare.
CONSNU
000000
number
Consecutive numbering. Allows sequential numbering for print operations. The consecutive
number value is incremented following each print operation.
The initial value of this parameter is set to the start up value specified on the CONSTU
parameter. Changing either CONSTU or CONSNU immediately resets the consecutive number
used for printing.
CONSTU
000000
number
Consecutive number start up value in the range of 000000–999999. Specifies the initial
consecutive number (CONSNU) value used when the indicator is powered on.
Table 3-6. Program Menu Parameters
XXXXXXX
XXXXXXX
PWRUPM
GO
OIML
REGULA
000000
number
CONSNU
DELAY
PROGRM PFORMT
SERIAL
CALIBR
CONFIG FORMAT
000000
number
CONSTU
NTEP
CANADA
COUNT
VERS
DIGIN ALGOUT
DA TFMT
MMDDYY
DA TE
DDMMYY
YYMMDD
YYDDMM
TIMFMT
12HOUR
TIME
24HOUR
TIMSEP
COLON
COMMA
ACCESS
SPLSIZ
10
20
50
100
PCWGT
DSABLE
ENABLE
NONE
number
000000
STNDBY
0-255
000
ACCUM
ON
OFF
DA TSEP
SLASH
DASH
SEMI
UID
5