Salter Brecknell SBI140 User Manual
Page 31

26
ON/OFF/
ZERO
+
UNIT
Normal weighing mode
(more than 3s)
Enter setup mode when sealed
calibration switch is on?
ON/OFF/
ZERO
+
TARE
Normal weighing mode
(more than 3s)
Enter calibration mode when sealed
calibration switch is on?
Ⅺ.
Parameters setting summary:
Para-
meter
x/xy
Factory
Set
Setting
Final
Set
P1.xy
00-15
05
Auto-off time: no auto-off; 01-15
minutes auto-off time
P2.x
0, 1, 2
2
0=only Hold function; 1=only Print
function; 2=Print and Hold function,
P3.xy
0,1,2,
3-50
2
0=no hold function; 1=hold lager weight
reading; 2=auto release hold function
when weight is below 10d and auto-hold
new stable weight (more than 10d);
3-50=unchangeable reading when the
variety is within +3~50d
P4.x
0,1,2,3,4,
5,6,7
2
0=no RS232 function; 1=output display
data when PRINT pressed; 2=output
gross, tare and net weight
when PRINT
pressed; 3=continuously output display
data; 4= continuously output gross, tare
and net weight; 5=output display data
one time when scale is stable; 6=output
gross, tare and net weight one time when
scale become stable; 7=Bio-RS232
compatible (8 data) to NCI-SP1;
P5.x
0, 1, 2, 3,
4
3
Baud rate for RS232: 0=1200bps,
1=2400bps,
2=4800bps,
3=9600bps
,
4=19200bps
P6.x
0, 1, 2
0
RS232 format: 0=8N1
, 1=7O1, 2=7E1