6

(2)

Choose the weight inner code or input working voltage to be

displayed. The working voltage range is 4.8V-8V. If the voltage

is not within range, it may damage the indicator. Voltage value

is shown like this “U x.xx” and its unit is V.

(3)

When A/D code of weight is shown, press this button to choose

filtered or un-filtered weight A/D data; when  is on, the data

is filtered.

(4)

Press this button to exit this mode, and auto-reset the indicator,

display all segments of LCD, full capacity… just like power on

again and then it goes back to normal weighing mode.

3.5 Calibration mode

HOLD

UNITS TARE

OFF/ON

(1)

Rotate the flashed position from right to left

(2)

Change the digit on flashed position. The digit can be changed

to 0, 1, 2…9; and be flashed.

(3)

Confirm receiving input data and go into next step.

(4)

Exit from calibration mode to normal work mode


Before calibrate the scale, you should prepare a standard weight (more than

10% of FS weight, and the unit is same as P10 setting) for calibration.

.

CALIBRATION:

1. Move away any weight on scale. When normal weighing mode, press and

hold down TARE and ON/OFF/ZERO buttons to enter calibration mode.

2. When the indicator shows” CAL-?”, the scale is ready for calibration. Press

WT

VOL

FLT

EXT