17

x = 8 : +5d

Choose the zero tracking range as per the stability of weighing system,

accuracy and weight excursion. The normal setting is +0.5d ~ +1.5d.

5.17) P17.x: Data filter intensity

x = 0 very weak

x = 1 : weak

x = 2 : middle

x = 3 : strong

The larger the chosen digit is, data filter intensity is stronger, and the speed

of data updating is lower. The normal setting is 2.

5.18) P18.x: Range of weight stability checking

x = 0 : +0.5d

x = 1…9 : +1d --- +9d

If the variety of weight reading is within the setting range continuously for

several times, the scale is recognized as stable. The normal setting is +1d ~

+3d.

5.19) P19.x: Overload range that can be displayed (when weight is lager than

range, “ˉˉˉˉˉ”will be shown):

x = 0 : FS+0d

x = 1 : FS+9d

x = 2 : 101%FS

x = 3 : 102%FS

x = 4 : 105%FS

x = 5 : 110%FS

x = 6 : 120%FS

x = 7 : 150%FS

x = 8 : 200%FS

x = 9 : No limitation

5.20) P20.x: Backlight on-off mode can be displayed

x = 0 : Backlight is always off

x = 1 : Backlight is always on

x = 2 : Backlight is auto on and auto off. It is auto off after 10s when scale

goes to stable and has no key operation, and it is auto on when scale is

unstable or there’s some key operation.