17
x = 8 : +5d
Choose the zero tracking range as per the stability of weighing system,
accuracy and weight excursion. The normal setting is +0.5d ~ +1.5d.
5.17) P17.x: Data filter intensity
x = 0 very weak
x = 1 : weak
x = 2 : middle
x = 3 : strong
The larger the chosen digit is, data filter intensity is stronger, and the speed
of data updating is lower. The normal setting is 2.
5.18) P18.x: Range of weight stability checking
x = 0 : +0.5d
x = 1…9 : +1d --- +9d
If the variety of weight reading is within the setting range continuously for
several times, the scale is recognized as stable. The normal setting is +1d ~
+3d.
5.19) P19.x: Overload range that can be displayed (when weight is lager than
range, “ˉˉˉˉˉ”will be shown):
x = 0 : FS+0d
x = 1 : FS+9d
x = 2 : 101%FS
x = 3 : 102%FS
x = 4 : 105%FS
x = 5 : 110%FS
x = 6 : 120%FS
x = 7 : 150%FS
x = 8 : 200%FS
x = 9 : No limitation
5.20) P20.x: Backlight on-off mode can be displayed
x = 0 : Backlight is always off
x = 1 : Backlight is always on
x = 2 : Backlight is auto on and auto off. It is auto off after 10s when scale
goes to stable and has no key operation, and it is auto on when scale is
unstable or there’s some key operation.