8

can press ON/OFF/ZERO to exit the calibration mode. Or load 5%-100%FS

weight, and use PRINT/HOLD and UNIT keys to input the loaded weight.

If the input data is larger than 199999, it will show as “Hxxx” and “Lxxx”. If

the triangular symbol on the left bottom of LCD window appears, it means

that the digit being changed is the displayed most significant bit which can

only be 0/blank or 1. Press TARE key to confirm your setting and the

indicator will flash the input standard weight. Wait till the scale comes steady,

and input A/D data as per the standard weight.

The indicator will automatically go to next step, if the second point can be

calibrated correctly. If there’s an error occurred, the scale will display “CAL.

Er” and return back to step 5 for re-calibration.

7. When ‘CAL.P2’ is displayed, the scale will be calibrated on third calibration

point. When xxxxxx kg (or lb) is displayed (100% FS is default), you can

press ON/OFF/ZERO to exit the calibration mode or Place a standard

weight (must be in the range of 10%-100% FS, and equal or larger than that

for the second calibration point; this is also the range of your input number)

on the scale. Use PRINT/HOLD and UNIT key to input the standard

weight’s value. Use TARE key to confirm the standard weight and input

number are correct. If the calibration weight for third point is same with that

for second point and the calibration weight is more than 10%FS, input the

standard calibration weight same as second point calibration and press TARE

key to confirm the setting. The indicator will flash the input weight. If the

indicator get reasonable data (the input weight is correct, and the calibration

weight of third calibration is more than equal to the calibration weight of

second calibration), it will go to next step automatically. If there’s an error

occurred, the scale will display “CAL.Er” and return back to step 5 for

re-calibration.

8. When ‘CAL.P0’ is shown again, that means the scale will calibrate scale’s

zero-point again. Now, you can press ON/OFF/ZERO to exit the calibration

mode; or Move away any weight on the scale, press TARE key to confirm;

the displayed data will blink. If the indicator gets reasonable data, it will

calculate and store all parameters in EEPROM. And then it will auto-reset

and display all segments of LCD, full capacity… like power on again. If

there’s an error occurred in calibration, the scale will display “CAL.Er” and