Salter Brecknell SBI140 User Manual
Page 25
![background image](/manuals/634868/25/background.png)
20
(4) Command: T
Response: TARE function is activated, and then returns scale status. just like
pressing TARE button:
H1H2H3
If TARE function cannot be activated, it will return to current scale status.
(5) Command: U
Response: Changes units of measure and return scale status with new units,
just like pressing UNIT button. The new measure unit should be allowed to
use as per P11 setting.
H1H2H3
If the weight unit is lb:oz, U1U2= lb oz
(6) Command: L
Response: If Hold function can be activated, it will enable/disable hold (lock)
function, like the HOLD key is pressed, and returns scale status.
(7) Command: X
Response: power off the scale, just like press down the ON/OFF key to turn
off the scale.
(8) Command: all others
Response: Unrecognized command
2.5) Output status bit meaning:
Table5: The status bits definition:
Bit
Byte 1 (H
1
)
Byte 2 (H
2
)
Byte 3 (H
3
)
0
0=stable
0= not under capacity 01=normal work mode
10= hold work mode
00=not define
11= not define
1= not stable
1= under capacity
1
0= not at zero point 0= not over capacity
1= at zero point
1= over capacity
2
always 0
always 0
0= gross weight
1= net weight
3
0= eeprom OK
always 0
always 0
1= eeprom error
4
always 1
always 1
always 1
5
always 1
always 1
always 1
6
always 0
always 1
always 0
7
parity
Parity
parity