beautypg.com

Specifications – Canon GALVANO GC-251 User Manual

Page 6

background image

CANON Digital Galvano Scanner System

KP-1SM Series + GC-251

Users Manual 1.00

6

2. Specifications

2.1.

Galvano Motor (KP-1SM30, KP-1SM100)

Performance and shape

KP-1SM30

KP-1SM100

Conforming

beam diameter

Φ10 to φ20

Φ20 to φ30

Scan angle

±20 deg

±20 deg

Number of encoder

pulses

83328 pulses/rotation

83328 pulses/rotation

Command resolution

0.036 μrad 0.036

μrad

Torque constant

0.041 Nm/A

0.093 Nm/A

Weight 0.95kg

1.65kg

(Reference) Command resolution

The controller of this system divides one encoder cycle into 2048.

KP-1SM30,SM100 is controlled by the following command resolution:

83328 pulses × 2048 divisions = 170655744 pulses/rotation (360°)

360°/ 170655744 = 0.000002109 = 0.036urad

Environmental conditions

Operating temperature and humidity

23

10 , 80% RH or less (No condensation)

Storage temperature and humidity

–20 to 60ºC, 90% RH or less (No condensation)

Note: The above operating temperature and humidity conditions depend on the operating and heat

radiation conditions.