Operating procedure – Canon GALVANO GC-251 User Manual
Page 34
CANON Digital Galvano Scanner System
KP-1SM Series + GC-251
Users Manual 1.00
34
4. Operating Procedure
This chapter explains how to use the system.
Operations by [GALILEO] and by [RS-232C command] are explained together.
4.1. Controller
Activation
When the power (+5 V, +24 V) is turned on, the controller first does the following automatically:
Reads saved parameters from ROM
Homes to the origin
Corrects the encoder (high-speed oscillation of a fixed angle)
It takes about 15 seconds until the controller is activated.
With the default settings at shipping, the controller is activated with the internal clock.
The completion of activation can be confirmed as follows:
GALILEO
Note: At activation, the status temporarily changes to “DISCONNECT” and the ALARM lamp lights
(about 5 s). This is not an error. Wait for a while.
Changes to
‘CONNECT’
Lit