Operators Instruction
Programming the AFI4 with the FlexProgram
Open the FlexProgram
Select AFIx in the bottom and this screen will appear
Screen DTM setup
It is recommended to select
-
“Frame language”
Select measurement mode:
-
Conductivity or Concentration
Temperature unit °C or °F can also be selected.
Screen Identification
Enter
TAG number
Tag:
8 characters
Long tag: 30 characters
Date of 1
st
use
(year/month/date)
Description
(16 characters)
Message
(32 characters)
Screen Input (when conductivity is selected in DTM setup)
Setting for the 4 conductivity ranges. Selectable from drop
down menu with 14 defined ranges.
Factory setting will be
0...200 mS/cm
0… 20 mS/cm
0… 2 mS/cm
0…500 µS/cm
How to connect the external input R1 and R2 is defined
in front of the ranges. (0=Not Connected / 1=24VDC)
All settings can be individually programmed with the Baumer FlexProgram and programming unit FlexProgrammer 9701. All parameters, such as
conductivity range, concentration and output type, diagnostics, data logging, tag no., damping etc. can be configured using the FlexProgrammer
9701.
Integrated HELP-menus will give full instruction.
The programming entered to the AFIx transmitter will automatically
be uploaded to the DFON display through the UnitCom ribbon cable,
if Transmitter Value is selected in the programme or by the touch screen.
If Display Conversion is selected the DFON display can be programmed
individually, e.g. with other engineering units or ranges.
English
CombiLyz, AFI4