Operators Instruction: 11126179 00 Page 3 / 36

Operators Instruction

Set-up the AFI4

Typical temperature dependence

Media

% change per °C

Acids

1,0 … 1,6

Bases

1,8 … 2,2

Salts

2,2 … 3,0

Neutral water

2,0

AFIx

range

Conductivity in different media:

Conductivity

Media group

Media

55

nS/cm

Water

Ultra-pure water

1

µS/cm

Pure water

10

µS/cm

Process water

100 µS/cm

Food

Drinking water

Beer

1

mS/cm

Milk

Orange juice

10

mS/cm

Apple juice

100 mS/cm

Process

Phosphoric acid

Hydrochloric acid

Conductivity ranges (selectable)

0…500 µS/cm

0...1 mS/cm

0...10 mS/cm

0...100 mS/cm

0...1 S/cm

0...2 mS/cm

0...20 mS/cm

0...200 mS/cm

0...3 mS/cm

0...30 mS/cm

0...300 mS/cm

0...5 mS/cm 0...50 mS/cm

0...500 mS/cm

Definition:

1.000 µS/cm = 1.0 mS/cm

1.000 mS/cm = 1.0 S/cm

Concentration ranges (selectable)

Range R1

R2

Range

R1

R2

1

N.C.

N.C. 3

N.C.

24 VDC

2

24 VDC N.C. 4

24 VDC

24 VDC

To set the external input for range selection

White background

Green background

Red background

Error message and red background

Selectable display views

Conductivity

Concentration

Visual alerts and colours

Value w. all values

Media w. values

Bar graph w. values

Concentration value in %

same views available as for conductivity

Value w. TAG

Media w. TAG

Bar graph incl. temp.

NaOH (caustic soda) 0...15% by weight (0...90°C)

25...50% by weight (0...90°C)

HNO

3

(nitric acid)

0...25% by weight (0...80°C)

36...82% by weight (0...80°C)

1 × customer defined (30 point linearization)

Conductivity measurement is very temperature dependent.

The factory set-up is 2,0%/°C.

This means that the AFIx will automatically compensate with this fac-
tor. If the temperature dependence is known for the media this factor
must be set. However if the factor is unknown, it is recommended to
use an average from above table.

It is possible to adjust the temperature compensation factor from 0,0 …
5,0 %/°C.

Configure the temperature compensation by the touch screen:

Configure the range by the touch screen:

Configuration by the FlexProgram and FlexProgrammer please see
HELP in the FlexProgram.

Concentration output can also be selected by the touch screen or pro-
grammed by the FlexProgram.

Programming of the display colours, warnings and relays can be done
by the touch screen (see page 8 and 9) or by the FlexProgrammer.