Operators Instruction: 11126179 00 Page 3 / 36
Operators Instruction
Set-up the AFI4
Typical temperature dependence
Media
% change per °C
Acids
1,0 … 1,6
Bases
1,8 … 2,2
Salts
2,2 … 3,0
Neutral water
2,0
AFIx
range
Conductivity in different media:
Conductivity
Media group
Media
55
nS/cm
Water
Ultra-pure water
1
µS/cm
Pure water
10
µS/cm
Process water
100 µS/cm
Food
Drinking water
Beer
1
mS/cm
Milk
Orange juice
10
mS/cm
Apple juice
100 mS/cm
Process
Phosphoric acid
Hydrochloric acid
Conductivity ranges (selectable)
0…500 µS/cm
0...1 mS/cm
0...10 mS/cm
0...100 mS/cm
0...1 S/cm
0...2 mS/cm
0...20 mS/cm
0...200 mS/cm
0...3 mS/cm
0...30 mS/cm
0...300 mS/cm
0...5 mS/cm 0...50 mS/cm
0...500 mS/cm
Definition:
1.000 µS/cm = 1.0 mS/cm
1.000 mS/cm = 1.0 S/cm
Concentration ranges (selectable)
Range R1
R2
Range
R1
R2
1
N.C.
N.C. 3
N.C.
24 VDC
2
24 VDC N.C. 4
24 VDC
24 VDC
To set the external input for range selection
White background
Green background
Red background
Error message and red background
Selectable display views
Conductivity
Concentration
Visual alerts and colours
Value w. all values
Media w. values
Bar graph w. values
Concentration value in %
same views available as for conductivity
Value w. TAG
Media w. TAG
Bar graph incl. temp.
NaOH (caustic soda) 0...15% by weight (0...90°C)
25...50% by weight (0...90°C)
HNO
3
(nitric acid)
0...25% by weight (0...80°C)
36...82% by weight (0...80°C)
1 × customer defined (30 point linearization)
Conductivity measurement is very temperature dependent.
The factory set-up is 2,0%/°C.
This means that the AFIx will automatically compensate with this fac-
tor. If the temperature dependence is known for the media this factor
must be set. However if the factor is unknown, it is recommended to
use an average from above table.
It is possible to adjust the temperature compensation factor from 0,0 …
5,0 %/°C.
Configure the temperature compensation by the touch screen:
Configure the range by the touch screen:
Configuration by the FlexProgram and FlexProgrammer please see
HELP in the FlexProgram.
Concentration output can also be selected by the touch screen or pro-
grammed by the FlexProgram.
Programming of the display colours, warnings and relays can be done
by the touch screen (see page 8 and 9) or by the FlexProgrammer.
English
CombiLyz, AFI4