Description

LSP 05X Potentiometric Level Transmitter

The level measuring device LSP05X utilises the potentiometric meas-
urement principle and can be used in all medias that have a minimum
conductivity of 50 µS/cm.

From the low resistive measuring rod a high frequent current is fed
through the media to the tank wall. The voltage between the tip of the
rod and the tank wall is measured. This output voltage is proportional
to the tank fi lling level.

The measurement is unaffected by temperature and adhesive media.

The instrument is ideal for measurements in small wessels with tough,
pasty or strong adhesive media, such as ketchup, honey, and tooth-
paste. The integrated electronics provide a 4...20 mA output.

The LSP has automatic recognition of top/buttom mounting position.
Even angled installation is possible. In a non-conductive tank such as
a plastic tank a reference rod must be installed.

Please observe that a non-linear conductivity in the media will affect
the accuracy of the measurement.

A version with separate rod sensor and electronics is available for
applications where the ambient temperature at the measuring point
exceeds 60°C. Due to the high temperature limit the LSP050 is well
suitable for CIP and SIP processes.

The hygienic installation is guaranteed by using one of the hygienic
weld-in sleeves. (Please refer to the separate data sheet).

Safety instructions

This instrument is built and tested ac cord ing to the current EU-direc-
tives and packed in technically safe con di tion. In order to main tain this
condition and to ensure safe op er a tion, the user must follow the hints
and warnings given in this instruction.

During the installation the valid national rules have to be observed.
Ignoring the warnings may lead to severe per son al injury or substantial
damage to property.

The product must be op er at ed by trained staff. Correct and safe
operation of this equipment is de pend ent on proper transport, storage,
installation and op er a tion.

All electrical wiring must conform to local stand ards. In order to prevent
stray elec tri cal radiation, we rec om mend twisted and shielded input
cables, as also to keep power sup ply cables separated from the input
cables. The con nec tion must be made ac cord ing to the connecting
di a grams.

Before switching on the power supply take care that other equip ment
is not af fect ed. Ensure that the supply voltage and the conditions in the
environment comply with the spec i fi ca tion of the de vice.

Before switching off the supply voltage check the possible effects on
other equipment and the processing system.

WARNING

This product contains no replaceable parts.
In case of malfunction the product must be shipped to Baumer for
repair.

LSP051