beautypg.com

Dynalco SC-2124M Scanner/Monitor User Manual

Page 10

background image

10

2.2 Wiring

(Cont’d)



Connectors D and E are input terminals, each requiring p/n SC-CBLE25MM cable / terminal
block assembly.

Thermocouples can be connected to any input of either connector D or E.

For linear inputs (4 – 20 mA or 0 – 5 VDC) it is recommended to connect the wires to the
appropriate terminals for connector D. However, either connector D or E may be used. See
previous page for information on DIP switch settings for 4 – 20 mA inputs.

Any RTD inputs require the RTD module p/n SC-RTD, which includes the cable assembly. In
this case ,p/n SC-CBLE25MM will not be required. Note that (1) SC-RTD will handle up to 12
RTD inputs and can be connected to either connector D (Ch # 1 – 12) or connector E (Ch #
13 – 24).

Instructions for SC-RTD module configuration are in Appendix E.



Connector D (DB-25F)

This Connector provides the Analog Inputs for Channels 1 – 12.

Pinout:
Pin#

Pin

Name

Pin#

Pin

Name

1

Channel 1 (+)

14

Channel 1 (-)

2

Channel 2 (+)

15

Channel 2 (-)

3

Channel 3 (+)

16

Channel 3 (-)

4

Channel 4 (+)

17

Channel 4 (-)

5

Channel 5 (+)

18

Channel 5 (-)

6

Channel 6 (+)

19

Channel 6 (-)

7

Channel 7 (+)

20

Channel 7 (-)

8

Channel 8 (+)

21

Channel 8 (-)

9

Channel 9 (+)

22

Channel 9 (-)

10

Channel 10 (+)

23

Channel 10 (-)

11

Channel 11 (+)

24

Channel 11 (-)

12

Channel 12 (+)

25

Channel 12 (-)